A2Nreference¶
SCPI Commands
READ:BLUetooth:SIGNaling<Instance>:RXQuality:IQDRange:LENergy:LE1M:A2NReference
FETCh:BLUetooth:SIGNaling<Instance>:RXQuality:IQDRange:LENergy:LE1M:A2NReference
CALCulate:BLUetooth:SIGNaling<Instance>:RXQuality:IQDRange:LENergy:LE1M:A2NReference
- class A2Nreference[source]
A2Nreference commands group definition. 3 total commands, 0 Sub-groups, 3 group commands
- class CalculateStruct[source]
Response structure. Fields:
Trans_Packets: float: decimal No. of transmitted packets Range: 0 to 200E+6
Rec_Events: float: No parameter help available
Rec_Events_Val_Crc: float: No parameter help available
No_Valid_Iqp_Airs: enums.ResultStatus2: decimal No. of valid measurements Range: 0 to 100E+3
Valid_Iq_Pairs_Rec: enums.ResultStatus2: float Percentage of received valid IQ sample pairs related to all received IQ pairs Range: 0 % to 100 % , Unit: %
Mean_Amplitude: enums.ResultStatus2: float Mean amplitude Range: 0 to 1 , Unit: dBFS
Max_Amplitude: enums.ResultStatus2: float Max amplitude Range: 0 to 1 , Unit: dBFS
Rec_Events_Cnt: enums.ResultStatus2: decimal No. of received events
Pro_Events_Cnt: enums.ResultStatus2: decimal No. of processed events
Pro_Events_Val_Crc: enums.ResultStatus2: No parameter help available
Nos_Pos_Amp: enums.ResultStatus2: No parameter help available
- class FetchStruct[source]
Response structure. Fields:
Trans_Packets: int: decimal No. of transmitted packets Range: 0 to 200E+6
Rec_Events: float: No parameter help available
Rec_Events_Val_Crc: int: No parameter help available
No_Valid_Iqp_Airs: int: decimal No. of valid measurements Range: 0 to 100E+3
Valid_Iq_Pairs_Rec: float: float Percentage of received valid IQ sample pairs related to all received IQ pairs Range: 0 % to 100 % , Unit: %
Mean_Amplitude: float: float Mean amplitude Range: 0 to 1 , Unit: dBFS
Max_Amplitude: float: float Max amplitude Range: 0 to 1 , Unit: dBFS
Rec_Events_Cnt: int: decimal No. of received events
Pro_Events_Cnt: int: decimal No. of processed events
Pro_Events_Val_Crc: int: No parameter help available
Nos_Pos_Amp: int: No parameter help available
- class ReadStruct[source]
Response structure. Fields:
Trans_Packets: float: decimal No. of transmitted packets Range: 0 to 200E+6
Rec_Events: float: No parameter help available
Rec_Events_Val_Crc: float: No parameter help available
No_Valid_Iqp_Airs: float: decimal No. of valid measurements Range: 0 to 100E+3
Valid_Iq_Pairs_Rec: float: float Percentage of received valid IQ sample pairs related to all received IQ pairs Range: 0 % to 100 % , Unit: %
Mean_Amplitude: float: float Mean amplitude Range: 0 to 1 , Unit: dBFS
Max_Amplitude: float: float Max amplitude Range: 0 to 1 , Unit: dBFS
Rec_Events_Cnt: int: decimal No. of received events
Pro_Events_Cnt: int: decimal No. of processed events
Pro_Events_Val_Crc: int: No parameter help available
Nof_Pos_Amp: int: decimal No. of possible amplitude measurements determined by CTE length, slot length and number of antennas
- calculate() → CalculateStruct[source]
# SCPI: CALCulate:BLUetooth:SIGNaling<Instance>:RXQuality:IQDRange:LENergy:LE1M:A2NReference value: CalculateStruct = driver.rxQuality.iqDrange.lowEnergy.le1M.a2Nreference.calculate()
Returns the results of IQ dynamic range Rx measurement for the specified antenna. Commands for uncoded LE 1M PHY (..:LE1M. .) and LE 2M PHY (..:LE2M..) are available. Commands for reference antenna (…:A0Reference) , mandatory second non-reference antenna (…:A1NReference) , and optional third and fourth non-reference antennas are available. The values described below are returned by FETCh and READ commands. CALCulate commands return limit check results instead, one value for each result listed below.
Use RsCmwBluetoothSig.reliability.last_value to read the updated reliability indicator.
- return
structure: for return value, see the help for CalculateStruct structure arguments.
- fetch() → FetchStruct[source]
# SCPI: FETCh:BLUetooth:SIGNaling<Instance>:RXQuality:IQDRange:LENergy:LE1M:A2NReference value: FetchStruct = driver.rxQuality.iqDrange.lowEnergy.le1M.a2Nreference.fetch()
Returns the results of IQ dynamic range Rx measurement for the specified antenna. Commands for uncoded LE 1M PHY (..:LE1M. .) and LE 2M PHY (..:LE2M..) are available. Commands for reference antenna (…:A0Reference) , mandatory second non-reference antenna (…:A1NReference) , and optional third and fourth non-reference antennas are available. The values described below are returned by FETCh and READ commands. CALCulate commands return limit check results instead, one value for each result listed below.
Use RsCmwBluetoothSig.reliability.last_value to read the updated reliability indicator.
- return
structure: for return value, see the help for FetchStruct structure arguments.
- read() → ReadStruct[source]
# SCPI: READ:BLUetooth:SIGNaling<Instance>:RXQuality:IQDRange:LENergy:LE1M:A2NReference value: ReadStruct = driver.rxQuality.iqDrange.lowEnergy.le1M.a2Nreference.read()
Returns the results of IQ dynamic range Rx measurement for the specified antenna. Commands for uncoded LE 1M PHY (..:LE1M. .) and LE 2M PHY (..:LE2M..) are available. Commands for reference antenna (…:A0Reference) , mandatory second non-reference antenna (…:A1NReference) , and optional third and fourth non-reference antennas are available. The values described below are returned by FETCh and READ commands. CALCulate commands return limit check results instead, one value for each result listed below.
Use RsCmwBluetoothSig.reliability.last_value to read the updated reliability indicator.
- return
structure: for return value, see the help for ReadStruct structure arguments.